Reported Jun 10, 2020 · Terminated
TRUE METRIX SELF MONITORING BLOOD GLUCOSE SYSTEM - Product Usage: is intended for the quantitative measurement of…
Trividia Health, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Trividia Health, Inc. · 2400 NW 55th Ct · Ft Lauderdale, FL · 33309-2672
Is this the product you have?
- TRUE METRIX SELF MONITORING
- BLOOD GLUCOSE SYSTEM - Product Usageis intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip or forearm
- Lot codes and dates
- Serial Number TA1548753
- Amount recalled
- 1 unit
Where it was sold
- Nationwide
- FL
Why it was recalled
One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL.