Reported Aug 6, 2025 · Ongoing
MR 7700 Model Numbers (REF): (1) 782120, (2) 782153;
Philips North America
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips North America · 222 Jacobs St · Cambridge, MA · 02141-2296
Is this the product you have?
- MR 7700 Model Numbers (REF)
(1) 782120(2) 782153
- Lot codes and dates
3 of 3 products
Products covered by this recall # Barcodes Product as published 1 Model Numbers (REF): (1) 782120, (2) 782153 2 0088483810411200884838112858UDI-DI: (1) 00884838104112, (2) 00884838112858 3 Serial Numbers: (1) 65000, 65001, 65002, 65003, 65004, 65005, 65006, 65007, 65008, 65009, 65011, 65012, 65013, 65014, 65015, 65016, 65017, 65018, 65019, 65020, 65021, 65023, 65024, 65025, 65026, 65027, 65028, 65029, 65030, 65031, 65032, 65034, 65035, 65037, 65038, 65039, 65040, 65041, 65042, 65043, 65049, 65050, 65051, 65056, 65057, 65059, 65060, 65061, 65063, 65064, 65065, 65066, 65067, (2) 68000, 68001, 68006, 68007, 68008, 68009, 68010, 68011, 68012, 68013, 68014, 68015, 68016, 68017, 68018, 68020, 68022, 68023, 68024, 68025, 68028, 68029, 68030, 68031, 68033, 68034, 68036, 68037, 68041 - Amount recalled
- 82 units
This recall covers 19 products
The agency filed one notice per product. You are reading one of them; the other 18 are below.
6 more are in the archive.
Where it was sold
- Nationwide
- GA
Why it was recalled
The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new produce recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.