Reported Jul 27, 2016 · Terminated

cobas¿ Liat¿ Analyzer

Roche Molecular Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Roche Molecular Systems, Inc. · 1080 US Highway 202 S · Branchburg, NJ · 08876-3733

Is this the product you have?

  • cobas¿ Liat¿ Analyzer
Lot codes and dates
  • US-IVDM1-E-00383 & M1-E-00434
Amount recalled
2

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.

This product could cause temporary or reversible health problems — stop using it.