Reported Sep 25, 2013 · Terminated

GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for…

GE Healthcare, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • GE Healthcare Discovery IGS 730 Cardiovascular X-Ray Imaging System. The angiography X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image guided surgical procedures. The OR table is suitable for interventional and surgical procedures.
Lot codes and dates
  • Mfg Lot or Serial # 00000611882BU2 00000612803BU7 00000613979BU4 00000620447BU3
Amount recalled
4 (2 US, 2 OUS)

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

Why it was recalled

GE Healthcare has recently become aware of a potential safety issue with respect to Discovery IGS 730 Imaging Systems. A patient data management error may prevent the Discovery IGS 730 Imaging System from booting on start-up or re-booting during procedure. The boot issue may occur when a large amount of patient data remains in the browser and is not deleted. The inability to boot the system ma

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 25, 2013 · Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.