Reported Jul 4, 2018 · Completed

ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model 10849000

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model 10849000
Lot codes and dates
    • 164033
    • 164071
    • 164034
    • 164095
    • 164057
    • 164055
    • 164054
    • 164080
    • 164072
    • 164088
    • 164048
    • 164094
    • 164068
    • 164019
    • 164013
    • 164073
    • 164027
    • 164026
    • 164025
    • 164060
    • 164083
    • 164021
    • 164066
    • 164075
    • 164084
    • 164086
    • 164074
    • 164098
    • 164028
    • 164053
    • 164063
    • 164064
    • 164024
    • 164049
    • 164022
    • 164018
    • 164100
    • 164097
    • 164011
Amount recalled
39

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Where it was sold

Why it was recalled

If the C-arm is manually moved in the longitudinal direction to a fully extended position, the system's safety stop mechanism may be activated. As a result, the C-arm stops the movement and can drop up to 20 mm (0.8 ). Under certain circumstances, the cover of the detector or collimator can touch and possibly injure the patient due to the C-arm drop. This could lead to a situation in which it is necessary to stop the clinical treatment or continue the treatment on an alternative system.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 18, 2020 · Data as of Sep 19, 4:45 PM UTC