Reported Aug 10, 2016 · Terminated
UroDiagnost; x-ray system. Product Code: 708033
Philips Electronics North America Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Electronics North America Corporation · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- UroDiagnost; x-ray system
- Product Code708033
- Lot codes and dates
- All serial numbers. 708033 System serial numbers 1, 17, 15, 4, 3, 7, 18, 21, 8, 13, 14, 11, 5, 16, 2.
- Amount recalled
- 993 total
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
The two bolts that attach the clamping plate to the lateral shaft may break, which could cause the c-arm to move laterally forward, slide out of its bearings, and sag
This product could cause temporary or reversible health problems — stop using it.