Reported Aug 10, 2016 · Terminated
AU680 Chemistry Analyzers, Reference No. B04779 The AU series are automated, clinical chemistry analyzers that measure…
Beckman Coulter Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232
Is this the product you have?
- AU680 Chemistry Analyzers, Reference No. B04779 The AU series are automated, clinical chemistry analyzers that measure analytes in patient samples (serum, plasma, urine, CSF) in combination with appropriate reagents, calibrators, quality, control (QC) materials, and other accessories.
- Lot codes and dates
- 1030834
- Amount recalled
- 3,084 units total (713 units in US)
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can leak onto the connectors of the Printed Circuit Boards (PCB) in the analyzer, resulting in a short-circuit causing the instrument to go into a stop mode.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.