Reported Sep 11, 2019 · Terminated
Norm-O-Temp model 111Z
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cincinnati Sub-Zero Products LLC, a Gentherm Company · 12011 Mosteller Rd · Cincinnati, OH · 45241-1528
Is this the product you have?
- Norm-O-Temp model 111Z
- Lot codes and dates
- Serial Number Range113-0001N-192-0798N
- UDI10613031861549 and 10613031861532 Part # 86153, 86154
- Amount recalled
- 674
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.