Reported Sep 11, 2019 · Terminated
Norm-O-Temp model 111W
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cincinnati Sub-Zero Products LLC, a Gentherm Company · 12011 Mosteller Rd · Cincinnati, OH · 45241-1528
Is this the product you have?
- Norm-O-Temp model 111W
- Lot codes and dates
- Serial number Range091-N6669-192-N7901
- UDI
10613031861563106130318615701061303186158710613031861617 Part # 86156861578615886161
- Amount recalled
- 1063
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Cincinnati Sub-Zero Products LLC, a Gentherm… recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.