Reported Sep 10, 2014 · Terminated

Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

  • Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.
Lot codes and dates
  • EasyLink Informatics System V4.0/V5 Siemens Material Number (SMN) 10444799, 10444800, 10487219
Amount recalled
1347

Where it was sold

  • Nationwide

Why it was recalled

May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 2, 2014 · Data as of Sep 19, 5:32 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.