Reported Oct 1, 2014 · Terminated

AXIOM Artis or Artis zee system Product Usage: Axiom Artis and Artis zee / zeego are dedicated angiography systems…

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 51 Valley Stream Pkwy · Malvern, PA · 19355

Is this the product you have?

  • AXIOM
  • Artis or Artis zee system Product UsageAxiom Artis and Artis zee / zeego are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures
Lot codes and dates
  • material number 3771750
  • model numbers
    • 5904441
    • 5904466
    • 7008605
    • 7412807
    • 7413078
    • 7555357
    • 7555365
    • 7555373
    • 7727717
    • 7728392
    • 10094135
    • 10094137
    • 10094139
    • 10094141
    • 10094142
    • 10094143
    • and 10280959
Amount recalled
1067

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

There is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 6, 2015 · Data as of Sep 19, 4:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.