Reported Oct 6, 2021 · Terminated
AMX Navigate Mobile X-Ray System
GE Healthcare, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- AMX Navigate Mobile X-Ray System
- Lot codes and dates
AMXNV2000001WKAMXNV2000002WKDF2401600003WK
- Amount recalled
- 3 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Why it was recalled
An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Healthcare, LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.