Reported Jul 8, 2026 · Ongoing

Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729…

Insulet Corporation

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Insulet Corporation · 50 & 100 Nagog Park · Acton, MA · 01720-3440

Is this the product you have?

6 of 6 products

Products covered by this recall
#Product as published
1POD-OMNI-I1-6720.
2POD-OMNI-I1-6220.
3OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod).
4OMNI-I1-6220.
5OMNI-I1-6720.
6BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).
Lot codes and dates

6 of 6 products

Products covered by this recall
#BarcodesProduct as published
12038508300057910385083000527Model/REF Number: POD-OMNI-I1-6720. Pack UDI-DI: 20385083000579. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01032521, PH1U01102521, PH1U12102421, PH1U12172421, PH1U11212421, PH1U11252421, PH1U11222421, PH1U12092421, PH1U09112421, PH1U11262421, PH1U04032521, PH1U12022421, PH1U12182421, PH1U04302524, PH1U05022521, PH1U01222521, PR1U11202533, PH1U09122421, PH1U10102421, PH1U12122421, PH1U01302522, PH1U02222521, PH1U05052521, PH1U10042421, PH1U12102412, PH1U01312521, PH1U03312521, PH1U05012521, PH1U09122422, PH1U09162421, PH1U09192421, PH1U09212421, PH1U09262411, PH1U09262421, PH1U10112422, PH1U10122421, PH1U10172421, PH1U12132422, PH1U01132521, PH1U01172521, PH1U02132511, PH1U03202521, PH1U03262521, PH1U04042521, PR1U11192531, PH1U11192511, PR1U01162621, PH1U09172421, PH1U09182421, PH1U09202421, PH1U10142421, PH1U10152421, PH1U10162421, PH1U01232522, PH1U01312511, PH1U02032521, PH1U02042521, PH1U02212521, PH1U03282521, PH1U04092521, PH1U04102521, PH1U04112521, PH1U10172522, PH1U10182521, PH1U10202521, PH1U10212521, PH1U10222521, PH1U10232521, PH1U10242521, PH1U10252521, PH1U11152521, PH1U11172511, PH1U11172521, PH1U11182521, PR1U11182531, PR1U11182541, PR1U11192541, PH1U11192521, PR1U11192532, PR1U11192542, PR1U11202521, PR1U11202532, PR1U11202541, PR1U11202542, PR1U11212521, PR1U11222531, PR1U11222541, PR1U11222521, PR1U11232541, PR1U11242521, PR1U11242531, PR1U11252541, PR1U11252512, PR1U11252521, PR1U11262521, PR1U12012521, PR1U12022521, PR1U12032521, PR1U12102522, PR1U12112521, PR1U12122521, PR1U12132511, PR1U12132521, PR1U12152521, PR1U12162521, PR1U12172521, PR1U01132621, PR1U01152621, PR1U01172621, PR1U01202621, PR1U01212621, PR1U01222621, PR1U01232621, PR1U01272621, PR1U01302641, PR1U01312631, PR1U02022641, PR1U02032631, PR1U02032641, PR1U02042641, PR1U02052631, PR1U02052641, PR1U02062631, PR1U02112641, PR1U02132611, PR1U02132641, PR1U11252531, PR1U11262541, PH1U06282422.
22038508300042510385083000411Model/REF Number: POD-OMNI-I1-6220. Pack UDI-DI: 20385083000425. Single Pod UDI-DI: 10385083000411. Lot Numbers: PH1U10182422, PH1U02202521, PH1U04222521, PH1U04232511, PH1U04232521, PH1U11142521, PH1U12042521, PH1U12052521, PH1U01092621.
3Model/REF Number: OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: PH1U01222522.
4203850830004310385083000411Model/REF Number: OMNI-I1-6220. Pack UDI-DI: 2038508300043. Single Pod UDI-DI: 10385083000411. Lot Numbers: PH1U04182521, PH1U04212521, PH1U12192411.
52038508300054810385083000527Model/REF Number: OMNI-I1-6720. Pack UDI-DI: 20385083000548. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01022521, PH1U12202421, PH1U12192421, PH1U10212421, PH1U10232421.
62038508300013510385083000114Model/REF Number: BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod). Pack UDI-DI: 20385083000135. Single Pod UDI-DI: 10385083000114. Lot Numbers: PH1U01072521, PH1U01242521, PH1U03082521, PH1U04172521, PH1U01272513.
Amount recalled
4,936,056 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Insulet Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:19 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.