Reported Jul 19, 2017 · Terminated
Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an…
St. Jude Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: St. Jude Medical, Inc. · 6901 Preston Rd · Plano, TX · 75024-2508
Is this the product you have?
- Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO
- Product UsageThe Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs
- Lot codes and dates
443 of 443 codes
AAD255.1AAN997.1AAP279.1AAP328.1AAP333.1AAT509.1AAT579.1AAT582.1AAT652.1AAT783.1AAT792.1AAT795.1AAT797.1AAT804.1AAT845.1AAT902.1AAT976.1AAT987.1AAT989.1AAU003.1AAU009.1AAU022.1AAU046.1AAU078.1AAU094.1ABA946.1ABM329.1ABM630.1ABT798.1ABT841.1ACE689.1ACE771.1ACE786.1ACE838.1ACE875.1ACE900.1ACE936.1ACE939.1ACE994.1ACF065.1ACK145.1ACK161.1ACK241.1ACK424.1ACK511.1ACK514.1ACK549.1ACK559.1ACK622.1ACK629.1ACK637.1ACK699.1ACK702.1ACK739.1ACK752.1ACK803.1ACK805.1ACK806.1ACK828.1ACK855.1
- Amount recalled
- 443 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
Why it was recalled
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new St. Jude Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.