Reported Jul 19, 2017 · Terminated
Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic…
St. Jude Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: St. Jude Medical, Inc. · 6901 Preston Rd · Plano, TX · 75024-2508
Is this the product you have?
- Infinity 7, Implantable Pulse Generator, REF 6663, Rx only
- STERILE EO Product UsageThe Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs
- Lot codes and dates
AAD257.1AAD302.1AAP362.1AAP375.1AAP385.1ALX096.1ALX257.1ALX258.1ALY268.1ALY409.1ALY435.1ALY632.1ALY640.1ALY763.1ALY780.1ALY788.1ALZ034.1ALZ042.1ANT267.1ANT731.1ZV3966.1ZW0230.1ZW0376.1ZW0429.1ZW0458.1ZW0498.1ZW0544.1ZW0545.1ZW0567.1ZW0570.1ZW0576.1ZW0590.1ZX5511.1ZY9379.1ZY9394.1ZY9398.1ZY9405.1ZY9413.1ZY9433.1ZY9530.1
- Amount recalled
- 40 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
Why it was recalled
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new St. Jude Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.