Reported Aug 5, 2020 · Terminated
StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid…
Medtronic Navigation, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic Navigation, Inc. · 826 Coal Creek Cir · Louisville, CO · 80027-9710
Is this the product you have?
- StealthStation" S7 System with Polaris Spectra Camera
- part number PSU 9733437 - Product Usageare intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures
- Lot codes and dates
- StealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852.
- Amount recalled
- 2
Where it was sold
Why it was recalled
There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic Navigation, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.