Reported Sep 14, 2016 · Terminated
NOMAD Pro Packaged X-Ray System, Part Number 08500009; NOMAD Pro Vet Packaged X-Ray System, Part Number 08500011; NOMAD…
Aribex Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Aribex Inc · 11727 Fruehauf Dr · Charlotte, NC · 28273-6507
Is this the product you have?
5 of 5 products
| Model / number | Product as published |
|---|---|
08500009 | NOMAD Pro Packaged X-Ray System |
08500011 | NOMAD Pro Vet Packaged X-Ray System |
08500013 | NOMAD Pro Packaged X-Ray System - Europe |
08500017 | NOMAD Pro X-Ray System, Remote Config. |
08500019 | NOMAD Pro Packaged X-Ray System JP |
- Amount recalled
- 16,871
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Aribex Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.