Reported Sep 14, 2016 · Terminated

NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022;…

Aribex Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Aribex Inc · 11727 Fruehauf Dr · Charlotte, NC · 28273-6507

Is this the product you have?

8 of 8 products

Products covered by this recall
Model / numberProduct as published
08500021Serial Number Range XD-100921-05 - XD-150109-11, 10000 - 20811, XD-140929-39 - XD-150123-20
08500022Serial Number Range XE-130528-03 - XE-150123-38, 10000 - 20541, XE-140908-32 - XE-150123-40
08500023Serial Number Range XF-121008-05 - XF-150205-03, 10003 - 20032, XE-150205-01
08500024Serial Number Range XG-130520-01 - XG-141210-05, 1000 - 20019
08500025Serial Number Range XI-140303-02 - XI-150210-9, XI-150202-04 - XI-150210-19
08500026Serial Number Range 10039 - 20041
08500028Part Number 08500027 - Serial Number Range 1000 - 20006, - Serial Number Range 10000 - 20030
08500029Serial Number Range 10000.
Amount recalled
16,871

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Aribex Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 26, 2017 · Data as of Sep 19, 5:20 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.