Reported Sep 14, 2016 · Terminated
BrightView XCT Upgrade Model 882454
Philips Medical Systems (Cleveland) Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131
Is this the product you have?
- BrightView XCT Upgrade Model 882454
- Lot codes and dates
110000211100008211000111
- Amount recalled
- 3 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.
This product could cause temporary or reversible health problems — stop using it.