Reported Oct 2, 2019 · Terminated
Atellica IM anti-CCP IgG (aCCP) Assay Siemens Material Number: 10732998 (100 Test) Test Version1.0 - Product Usage: The…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516
Is this the product you have?
- Atellica IM anti-CCP IgG (aCCP)
- Assay Siemens Material Number10732998 (100 Test) Test Version1.0 -
- Product UsageThe ADVIA Centaur¿ Anti-CCP IgG (aCCP) assay is for in vitro diagnostic use in the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma (K2-EDTA, lithium heparin) using the ADVIA Centaur XP system. Detection of anti-CCP antibodies is used as an aid in the diagnosis of rheumatoid arthritis (RA) and should be used in conjunction with other clinical information. Autoantibody levels represent one parameter in a multi-criteria diagnostic process, encompassing both clinical and laboratory-based assessments. [Note: the Atellica IM anti-CCP IgG (aCCP) assay was introduced for sale through the Replacement Reagent Policy.]
- Lot codes and dates
- All lots
- Amount recalled
- 169 kits
Where it was sold
Why it was recalled
Atellica IM anti-CCP IgG (aCCP) Test Definition (TDef) Mitigation Failure-reagent carryover mitigation for the Testosterone II (TSTII) assay is unable to be completed as defined in the aCCP Test Definition (TDef). The issue occurs on an Atellica IM Analyzer and is isolated to TDef Version 1.0. FOR OUS: Atellica IM anti-CCP IgG (aCCP) and Pregnancy-Associated Plasma Protein-A (PAPP-A) Test Definition (TDef) Mitigation Failure
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.