Reported Aug 12, 2026 · Ongoing
REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan…
Beckman Coulter Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232
Is this the product you have?
REF # D23354MicroScan LabPro V5.03 Panel Update 11 DVD
- Kit (US) on the following instruments- B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
- Lot codes and dates
- UDI15099590799250/ V5.03 Panel Update 11
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Beckman Coulter Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.