Reported Sep 9, 2020 · Terminated

Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool…

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355

Is this the product you have?

  • Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.
Lot codes and dates
  • Serial Numbers

    111 of 111 codes

    • 100413
    • 100414
    • 100417
    • 101016
    • 101017
    • 101019
    • 101020
    • 101022
    • 101029
    • 101031
    • 101036
    • 100263
    • 100416
    • 102008
    • 101083
    • 102138
    • 102019
    • 102021
    • 102025
    • 102029
    • 102145
    • 102147
    • 100346
    • 100349
    • 102199
    • 100095
    • 100097
    • 100103
    • 100115
    • 100116
    • 100117
    • 100120
    • 100121
    • 102158
    • 102167
    • 102168
    • 102172
    • 102272
    • 102068
    • 102069
    • 102078
    • 102080
    • 102093
    • 102221
    • 100163
    • 100165
    • 100306
    • 100309
    • 101056
    • 101099
    • 101100
    • 101101
    • 101060
    • 101068
    • 101069
    • 101070
    • 101071
    • 101072
    • 100064
    • 100422
  • Expanded S/N11007641 11007641 11007641 ***9/28/20 Expanded S/N*** 103053 103040 103042 103008
Amount recalled
110 units US

Where it was sold

  • Nationwide

Why it was recalled

The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 2, 2021 · Data as of Sep 19, 6:53 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.