Reported Sep 30, 2020 · Terminated

NM/CT 870 DR Model # H3100AS

GE Healthcare, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • NM/CT 870 DR Model # H3100AS
Lot codes and dates
  • Serial # 870X64001
  • System ID030020NU06 Serial # 870X64004
  • System IDM4018555 Serial # 870X64012
  • System IDX634445001 Serial # 870X64007
  • System IDV2728113 Serial # 870X64009
  • System IDM4016255 Serial # 870X64026
  • System IDM40230109 Serial # 870X64017
  • System IDM40240121 Serial # 870X64018
  • System IDM40240121 Serial # 870X64015
  • System IDM40345223 Serial # 870X64016
  • System ID499221NU01 Serial # 870X64031
  • System ID083049209623319 Serial # 870X64014
  • System IDIL1028NM41 Serial # 870X64022
  • System IDA5173220 Serial # 870X64019
  • System ID0850260167XE6 Serial # 870X64020
  • System ID0850260169 Serial # 870X64025
  • System IDEMM0135 Serial # 870X64010
  • System IDNM2882PA Serial # 870X64027
  • System IDPL0398NM01 Serial # 870X64029
  • System IDPL1031NM01 Serial # 870X64030
  • System IDPL2190NM01 Serial # 870X64028
  • System IDPL0372NM01 Serial # 870X64023
  • System ID600042NM10 Serial # 870X64013
  • System IDNP473631 Serial # 870X64021
  • System IDG001NU12 Serial # 870X64008
  • System ID00669NUC03
Amount recalled
31 units

This recall covers 8 products

The agency filed one notice per product. You are reading one of them; the other 7 are below.

Where it was sold

Why it was recalled

Rotor bearing screws were found loose on detector.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 6, 2021 · Data as of Sep 19, 6:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.