Medtronic Set Supplemental INS AxiEM ENT, Ref. 9734378, reusable, RX. This kit contains 9734308, Small Straight…
Medtronic Navigation, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic Navigation, Inc. · 826 Coal Creek Cir · Louisville, CO · 80027-9710
Is this the product you have?
- Medtronic Set Supplemental INS AxiEM ENT, Ref. 9734378, reusable, RX. This kit contains 9734308, Small Straight suction. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
- Lot codes and dates
- Each of the devices within the kit are labeled with information specific to each of the reusable devices within the kit and those devices are what goes into the customer inventory. The outer box where the lot number is printed is discarded by the customer on receipt and the separate devices within the box are what is stored on the shelf. Therefore, no kit lot number is being provided.
- Amount recalled
- 8 kits
This recall covers 8 products
The agency filed one notice per product. You are reading one of them; the other 7 are below.
Where it was sold
- Nationwide
Why it was recalled
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic Navigation, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.