Reported Sep 25, 2024 · Ongoing

Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005

Is this the product you have?

  • Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers
Lot codes and dates
  • UDI-DI00630414596310
  • Lot Numbers
    • 232033
    • 232128
    • 232137
    • 232146
    • 232147
    • 242149
    • 242150
    • 242365
Amount recalled
18483 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • GA

Why it was recalled

Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.