In the agency's Nov 11, 2026 enforcement report · Ongoing
TruSystem 7500 surgical table, Model/Catalog Number: 4091000
Baxter Healthcare Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Baxter Healthcare Corporation · 1 Baxter Pkwy · Deerfield, IL · 60015-4625
Is this the product you have?
- TruSystem 7500 surgical table
- Model/Catalog Number4091000
- Lot codes and dates
- Serial numbers
129 of 129 codes
108491706108726496108758657108858029108945825109083212109343910109684319110016886108557062108736192108773032108882327108976094109174333109370104109690845110016887108574826108736193108773033108882328108976095109174334109396256109690846110072583108578755108736194108842519108883564108976101109174335109464647109708813110084982108580717108736195108842523108883565109047445109174336109464648109708816110086138108580718108743656108842524108896687109047446109228237109464649109885369110144223108580720108743657108842525108896688109047447109228238
- UDI-DI number00887761968639
- Serial numbers
- Amount recalled
- 129 units
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Baxter Healthcare Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.