Reported Jul 1, 2026 · Ongoing · Updated Sep 16, 2026

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack…

Dexcom, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Dexcom, Inc. · 6340 Sequence Dr · San Diego, CA · 92121-4356

Is this the product you have?

  • Brand NameDexcom G7 Continuous
  • Glucose Monitoring System Product NameDexcom G7 US Retail Sensor/Transmitter 1-pack
  • Model/Catalog NumberSTP-AT-012
  • Software VersionN/A
  • Product DescriptionDexcom G7 US Retail
  • Sensor/Transmitter 1-pack ComponentDexcom G7 Continuous Glucose Monitoring System
Lot codes and dates
  • Lot CodeUDI-DI: 00386270004109
  • Lot Numbers1725204004 and 1725069002

Why it was recalled

Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Dexcom, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice

Data as of Sep 19, 6:28 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.