Reported Sep 2, 2026 · Ongoing · Updated Sep 16, 2026

PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209

GE Medical Systems Information Technologies Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Medical Systems Information Technologies Inc · 3114 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

    • PORTRAIT CORE SERVICES SOFTWARE V1.1
    • Ref# 5700209
Lot codes and dates
  • GTIN 00195278679284
  • Serial Numbers
    • SWT25250001TA
    • SWT25360001TA
    • SWT25360002TA
    • SWT24090006TA
    • SWT25100001TA
    • SWT25120002TA
    • SWT24140007TA
    • SWT24150008TA
    • SWT25100004TA
    • SWT24050001TA
    • SWT25110003TA
    • SWT24180011TA
    • SWT25120003TA
    • SWT25230001TA
    • SWT25230002TA
    • SWT25110009TA
    • SWT25110010TA
    • SWT25230003TA
    • SWT25230005TA
    • SWT25230006TA
    • SWT25230004TA
    • SWT25260001TA
    • SWT31240012TA
Amount recalled
23 units

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

Why it was recalled

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Medical Systems Information Technologies Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (Enforcement Report)

Read the FDA (weekly enforcement report) notice

Data as of Sep 19, 3:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.