Reported Sep 23, 2026 · Ongoing
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800…
GE Medical Systems Information Technologies Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Medical Systems Information Technologies Inc · 3114 N Grandview Blvd · Waukesha, WI · 53188-1615
Is this the product you have?
- Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
- Lot codes and dates
- 5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV
- Amount recalled
- 29,570 (23,632 US; 5,938 OUS)
Where it was sold
- Nationwide
Why it was recalled
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Medical Systems Information Technologies Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.