Reported Nov 5, 2014 · Terminated

Proteus 235 Proton Therapy System for cancer treatment.

Ion Beam Applications S.A.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ion Beam Applications S.A. · Chemin du Cyclotron, 3 · Louvain La Neuve, N/A · N/A

Is this the product you have?

  • Proteus 235 Proton Therapy System for cancer treatment.
Lot codes and dates
    • PAT.107
    • PAT.109
    • PAT.112
    • PAT.113
    • PAT.114
    • PAT.115
    • PAT.116
    • PAT.119
Amount recalled
8

Where it was sold

Why it was recalled

While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new infant & baby recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 10, 2014 · Data as of Sep 19, 6:49 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.