Reported Nov 5, 2014 · Terminated
Proteus 235 Proton Therapy System for cancer treatment.
Ion Beam Applications S.A.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ion Beam Applications S.A. · Chemin du Cyclotron, 3 · Louvain La Neuve, N/A · N/A
Is this the product you have?
- Proteus 235 Proton Therapy System for cancer treatment.
- Lot codes and dates
PAT.107PAT.109PAT.112PAT.113PAT.114PAT.115PAT.116PAT.119
- Amount recalled
- 8
Where it was sold
Why it was recalled
While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to accelerate protons) may generate a hot spot on a pencil beam position within a tumor, generating a dose that might exceed the prescribed dose on a single voxel.