Reported Nov 11, 2015 · Terminated

HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is…

Hamilton Medical, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Hamilton Medical, Inc. · 4990 Energy Way · Reno, NV · 89502-4123

Is this the product you have?

  • HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31
  • AnesthesiologyThe HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients
Lot codes and dates
  • software versions between V2.00 and V2.31
  • Ventilator catalog number1590001
  • Software catalog number159700
  • Serial numbers

    1,043 of 1,043 codes

    • 1231
    • 1233
    • 1234
    • 1235
    • 1014
    • 1296
    • 1593
    • 1594
    • 1595
    • 1626
    • 1628
    • 1629
    • 1631
    • 1632
    • 1633
    • 1634
    • 1637
    • 1693
    • 1697
    • 1862
    • 1863
    • 2021
    • 2411
    • 2412
    • 2413
    • 2414
    • 2416
    • 2417
    • 2418
    • 2591
    • 2592
    • 2593
    • 2594
    • 2595
    • 2681
    • 2683
    • 2686
    • 2687
    • 2689
    • 2776
    • 2777
    • 2778
    • 2779
    • 2780
    • 2781
    • 2784
    • 2785
    • 2786
    • 2787
    • 2789
    • 2790
    • 2793
    • 2794
    • 2798
    • 2803
    • 2887
    • 2947
    • 2948
    • 2951
    • 2952
Amount recalled
1128

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Hamilton Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 24, 2017 · Data as of Sep 19, 4:44 PM UTC