Reported Sep 27, 2023 · Ongoing
Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator
Hamilton Medical, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hamilton Medical, Inc. · 4655 Aircenter Cir · Reno, NV · 89502-5948
Is this the product you have?
- Hamilton-C1
- REF161001, Portable Intensive Care Ventilator
- Lot codes and dates
- UDI07630002800747 & 07630002813426 /
- Affected Serial Numbers
164 of 164 codes
241772419624216242172421924247242502432424360243702443124460244672448024487245112452324552245562455724558245612456424565245712457224574245782458324587245912459524612246162462024622246512466224672246912496524991250062501025011250142502225096251092511125118251412514225151251522515825163251692517125178
- Amount recalled
- 164 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
Stop using this product now
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Hamilton Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.