Reported May 18, 2016 · Terminated
Hamilton G5 with software version between v2.41, v2.42, v2.50. Catalog number: 1590001 and 1590002. Anesthesiology:…
Hamilton Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hamilton Medical, Inc. · 4990 Energy Way · Reno, NV · 89502-4123
Is this the product you have?
- Hamilton G5 with software version between v2.41, v2.42, v2.50
- Catalog number1590001 and 1590002
- AnesthesiologyHamilton G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infants and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care
- Lot codes and dates
- Serial numbers
1,115 of 1,115 codes
127816391686168916911692169516971699186419842412241324142417246525052506252125222523252425252526252725302531253225912592259325942595277627802876287728782880288128822883288428852886289028943365336633683370337736113881388238843885388638873888
- Serial numbers
- Amount recalled
- 1115
Why it was recalled
After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Hamilton Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.