Reported Oct 22, 2025 · Ongoing
Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01…
Spacelabs Healthcare, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Spacelabs Healthcare, Inc. · 35301 Se Center St · Snoqualmie, WA · 98065-9216
Is this the product you have?
- Brand NameBranding is not applied to this product
- Product NameDVI
- display cables Model/Catalog Number012-0895-01
- Software VersionN/A
- Product Description6 foot DVI-I to DVI-I male/male display cable
- ComponentYes, the cable is an external display cable used with the 91390 and 91393
- Lot codes and dates
- Lot CodeUDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube)
- Amount recalled
- 8179
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
Why it was recalled
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Spacelabs Healthcare, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.