Reported Aug 17, 2022 · Ongoing

Xhibit Telemetry Receiver, Model: 96280

Spacelabs Healthcare, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Spacelabs Healthcare, Inc. · 35301 Se Center St · Snoqualmie, WA · 98065-9216

Is this the product you have?

  • Xhibit Telemetry Receiver
  • Model96280
Lot codes and dates
  • UDI-DI
    • 10841522100246
    • Software version 1.4.0
  • Serial Numbers

    529 of 529 codes

    • 6280-001538
    • 6280-000590
    • 6280-000589
    • 6280-000339
    • 9280-000711
    • 6280-001103
    • 6280-001102
    • 6280-001469
    • 6280-001467
    • 6280-001686
    • 6280-001399
    • 6280-001323
    • 6280-000457
    • 6280-000456
    • 6280-000455
    • 6280-000440
    • 6280-000439
    • 6280-000438
    • 6280-000429
    • 6280-000476
    • 6280-000475
    • 6280-001693
    • 6280-000703
    • 6280-000702
    • 6280-000701
    • 6280-000700
    • 6280-000314
    • 6280-000258
    • 6280-000157
    • 6280-001223
    • 6280-001222
    • 6280-001221
    • 6280-001220
    • 6280-001219
    • 6280-001218
    • 6280-001217
    • 6280-001216
    • 6280-001215
    • 6280-001214
    • 6280-001207
    • 6280-001127
    • 6280-001125
    • 6280-001124
    • 6280-001123
    • 6280-001056
    • 6280-001055
    • 6280-001054
    • 6280-001053
    • 6280-001052
    • 6280-001051
    • 6280-001050
    • 6280-001288
    • 6280-001287
    • 6280-001490
    • 6280-001692
    • 6280-000588
    • 6280-000517
    • 6280-000516
    • 6280-000515
    • 6280-001668
Amount recalled
529

Where it was sold

Why it was recalled

Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Spacelabs Healthcare, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 3:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.