Reported Apr 15, 2026 · Ongoing
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Spacelabs Healthcare, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Spacelabs Healthcare, Inc. · 35301 Se Center St · Snoqualmie, WA · 98065-9216
Is this the product you have?
- Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
- Lot codes and dates
- Serial Numbers
1,792 of 1,792 codes
UDI-DI: 1084152210641510841522106422. 1496-2237361496-2237371496-2237381496-2243751496-2243761496-2244001496-2244011496-2244021496-2244031496-2244041496-2244051496-2244061496-2238501496-2238511496-2238521496-2244481496-2249561496-2249571496-2249581496-2249591496-2249601496-2249611496-2249811496-2242581496-2235951496-2235961496-2235971496-2235981496-2236191496-2236201496-2236211496-2236221496-2236231496-2236241496-2247171496-2247181496-2238541496-2243731496-2243741496-2238941496-2238951496-2238961496-2238971496-2238981496-2238991496-2239001496-2239011496-2239021496-2239031496-2239041496-2239051496-2239061496-2239071496-2239181496-2250431496-2250441496-2250451496-2250461496-225047
- Serial Numbers
- Amount recalled
- 1790
Where it was sold
Why it was recalled
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Spacelabs Healthcare, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.