Reported Nov 19, 2014 · Terminated
The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal…
Steris Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Steris Corporation · 5960 Heisley Rd · Mentor, OH · 44060-1834
Is this the product you have?
- The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.
- Lot codes and dates
- 5085 and 5085SRT
- Amount recalled
- 999
Where it was sold
Why it was recalled
The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
This product could cause temporary or reversible health problems — stop using it.