Reported Nov 13, 2019 · Terminated
Hitachi Scenaria Whole-body X-ray CT System
Hitachi Medical Systems America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hitachi Medical Systems America Inc · 1959 Summit Commerce Park · Twinsburg, OH · 44087-2371
Is this the product you have?
- Hitachi Scenaria Whole-body X-ray CT System
- Lot codes and dates
Scenaria units S5002-S5014S5017-S5044
- Amount recalled
- 39
Where it was sold
Why it was recalled
There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
This product could cause temporary or reversible health problems — stop using it.