Hitachi Scenaria CT system Product Usage: The SCENARIA CT system is indicated for head, whole body, cardiac and…
Hitachi Medical Systems America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hitachi Medical Systems America Inc · 1959 Summit Commerce Park · Twinsburg, OH · 44087-2371
Is this the product you have?
- Hitachi
- Scenaria CT system Product UsageThe SCENARIA CT system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the SCENARIA can be post processed by the SCENARIA to provide additional information. Post processing capabilities included in the SCENARIA software include CT angiography (CTA), Multiplanar reconstruction (MPR) and volume rendering. Volume datasets acquired by the SCENARIA Phase 3 can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician
- Lot codes and dates
- S7004
- Amount recalled
- 1
Where it was sold
Why it was recalled
The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Hitachi Medical Systems America Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.