Reported Feb 26, 2020 · Terminated
UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic…
Hitachi Medical Systems America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hitachi Medical Systems America Inc · 1959 Summit Commerce Park · Twinsburg, OH · 44087-2371
Is this the product you have?
- UST-2265-2 Probe
- Part NumberUST-2265-2 -
- Product UsageThe UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel
- Lot codes and dates
- Serial Number
G3007689G3007688G3007687206H2628206H2629206H2630206H2620206H2621206H2622206H2623206H26224206H2625206H2626206H2627205X3897205X3898205X3899and 205X3900
- Serial Number
- Amount recalled
- 18
Where it was sold
Why it was recalled
These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Hitachi Medical Systems America Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.