Reported Jan 8, 2014 · Terminated
SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E…
Steris Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Steris Corporation · 5960 Heisley Rd · Mentor, OH · 44060-1834
Is this the product you have?
- SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
- Lot codes and dates
- Model #6500 & 7500
- Amount recalled
- 6,028 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Steris Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.