Reported Mar 6, 2019 · Terminated
MR-OR 1.5T, Model Number 781437. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems Nederlands · Veenpluis 4-6 · Best · N/A
Is this the product you have?
- MR-OR 1.5T, Model Number 781437. Nuclear Magnetic Resonance Imaging System
- Lot codes and dates
- All systems
- Amount recalled
- 11226 total
This recall covers 64 products
The agency filed one notice per product. You are reading one of them; the other 63 are below.
51 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
This product could cause temporary or reversible health problems — stop using it.