Reported Apr 27, 2022 · Ongoing

Artis zee multi-purpose, Model Number 10094139

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Artis zee multi-purpose, Model Number 10094139
Lot codes and dates
  • UDI4056869010076
  • Serial Numbers

    264 of 264 codes

    • 158469
    • 157509
    • 158449
    • 158467
    • 157660
    • 157458
    • 130110
    • 158474
    • 130028
    • 130017
    • 158170
    • 158650
    • 158652
    • 158742
    • 158624
    • 158583
    • 157698
    • 130252
    • 158530
    • 157532
    • 130241
    • 157449
    • 158413
    • 157657
    • 157136
    • 130109
    • 157988
    • 157690
    • 158091
    • 158663
    • 159034
    • 158225
    • 157560
    • 130210
    • 157436
    • 157677
    • 157528
    • 159038
    • 158422
    • 157615
    • 130240
    • 157448
    • 157165
    • 158552
    • 158556
    • 158557
    • 158561
    • 158697
    • 158701
    • 159032
    • 157977
    • 157602
    • 159046
    • 157168
    • 130212
    • 130224
    • 158605
    • 158183
    • 130253
    • 158608
  • Additional Serial Numbers as of 10/4/23
    • 158138
    • 159065
    • 159085
    • 159079
    • 159080
    • 159088
    • 157674
    • 158113
    • 159030
    • 157659
    • 158136
    • 157668
    • 158079
Amount recalled
277 systems

This recall covers 18 products

The agency filed one notice per product. You are reading one of them; the other 17 are below.

5 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:20 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.