Reported Apr 27, 2022 · Ongoing

Artis Q biplane, Model Number 10848282

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Artis Q biplane, Model Number 10848282
Lot codes and dates
  • UDI4056869009995
  • Serial Numbers

    262 of 262 codes

    • 134007
    • 121314
    • 134008
    • 121396
    • 121590
    • 121665
    • 121564
    • 122025
    • 121627
    • 121185
    • 121430
    • 121406
    • 121336
    • 122012
    • 121471
    • 121195
    • 121467
    • 121612
    • 122022
    • 121530
    • 121089
    • 121094
    • 134004
    • 121169
    • 121006
    • 122021
    • 122028
    • 121013
    • 121040
    • 121501
    • 121368
    • 121526
    • 121208
    • 121548
    • 121515
    • 121445
    • 121446
    • 121623
    • 121082
    • 121459
    • 121011
    • 122037
    • 121039
    • 121521
    • 121371
    • 122005
    • 121374
    • 121204
    • 121614
    • 121305
    • 121536
    • 121053
    • 121054
    • 121422
    • 121014
    • 121112
    • 121325
    • 121357
    • 121621
    • 121167
  • Additional Serial Numbers as of 10/4/23134502 121340 121461 121051 121179 121016 121022
Amount recalled
269 systems

This recall covers 18 products

The agency filed one notice per product. You are reading one of them; the other 17 are below.

5 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.