Reported Apr 17, 2013 · Terminated
Medtronic MiniMed Paradigm Insulin Infusion Pumps Model Numbers: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715,…
Medtronic MiniMed
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic MiniMed · 18000 Devonshire St · Northridge, CA · 91325-1219
Is this the product you have?
- Medtronic MiniMed Paradigm
- Insulin Infusion Pumps Model Numbers
MMT-511MMT-512MMT-712MMT-712EMMT-515MMT-715MMT-522MMT-522KMMT-722MMT-722KMMT-523MMT-523KMMT-723MMT-723KMMT-554and MMT-754
- Product UsageParadigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes
- Lot codes and dates
- Not available
- Amount recalled
- 428,000 units
Where it was sold
Why it was recalled
Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become detached from the pump case and protrude from the lower right side of the pump. When the pump is expose to water it may result in damage to the pump's internal electronics. This moisture damage can prevent the pump's buttons from working properly or can cause the pump to alarm.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic MiniMed recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.