Reported Mar 3, 2021 · Terminated

OT1000 Series Orthopedic Surgical Tables

Steris Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Steris Corporation · 5960 Heisley Rd · Mentor, OH · 44060-1834

Is this the product you have?

  • OT1000 Series Orthopedic Surgical Tables
Lot codes and dates
    • Model No.: OT1000
    • OT1100
    • OT1200
Amount recalled
272 tables total (Domestic: 227 tables, Foreign 45 tables)

Where it was sold

Why it was recalled

On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Steris Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 10, 2021 · Data as of Sep 19, 7:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.