Reported Feb 26, 2020 · Terminated

ADVIA Chemistry Enzymatic Creatinine (ECRE_2) reagent, SMN 10335869

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005

Is this the product you have?

  • ADVIA Chemistry Enzymatic Creatinine (ECRE_2) reagent, SMN 10335869
Lot codes and dates
  • UDI Number00630414526690
  • LotsAll lots
Amount recalled
1,015 units

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Where it was sold

  • Nationwide

Why it was recalled

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 25, 2020 · Data as of Sep 19, 5:32 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.