ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005
Is this the product you have?
- ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP)
- Calibrator - product UsageFor in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay
- Lot codes and dates
- Siemens Material Number (SMN)ADVIA
- Chemistry10335897
- REF05006455 Lot# 453610
- Expiration Date2020-01
- UDI NumberADVIA: (01)00630414526843(10)453610(17)20200128
- Amount recalled
- 844
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.