Reported Jun 29, 2022 · Ongoing

REF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO,

Bard Peripheral Vascular Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Bard Peripheral Vascular Inc · 1625 W 3rd St Bldg 1 · Tempe, AZ · 85281-2438

Is this the product you have?

  • REF
    • PIG1260K
    • 6Fr
    • X10
    • Safe-t-Centesis Catheter drainage Kit
    • Sterile EO
Lot codes and dates
  • Lot # 0001405396
  • UDI/DI(01)10885403076114
Amount recalled
1460 units

Where it was sold

Why it was recalled

There is the potential that a 6 Fr Catheter Drainage kit contains an 8 Fr needle instead of a 6 Fr needle. They can be used in conjunction with the following drainage methods: Universal extension set and drainage bag, vacuum bottle, and wall suction.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Bard Peripheral Vascular Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:55 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.