Reported Mar 15, 2017 · Terminated

ADVIA Centaur Progesterone; Catalog Number: 10310305 10315522 10333111 Product Usage: The PROG method is an in vitro…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516

Is this the product you have?

  • ADVIA Centaur Progesterone
  • Catalog Number10310305 10315522 10333111
  • Product UsageThe PROG method is an in vitro diagnostic test for the quantitative measurement of progesterone in human serum and plasma on the Dimension Vista¿System. Progesterone measurements are used in the diagnosis and treatment of disorders of the ovaries or placenta
Lot codes and dates
  • SMN10310305:
    • 75422274
    • 83380274
    • 89389274
    • 99473274
    • 8713275
    • 19796275
    • 30565275
    • 38674275
    • 46515276
    • 55704276
    • 66971276
    • 78934276
    • 80930277
  • SMN10315522:
    • 74355274
    • 80476274
    • 87497274
    • 20274
    • 7772275
    • 14822275
    • 25943275
    • 35303275
    • 43229276
    • 46525276
    • 55505276
    • 67430276
    • 80448276
    • 81415277
  • SMN10333111: 84848274 89946274 14006275 33598275 49907276 66978276
Amount recalled
76,107

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

DHEA-S causes falsely elevated progesterone results.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 18, 2019 · Data as of Sep 19, 5:28 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.